Skip to main content

Table 5 Summary of agreement properties of ZB

From: Psychometric properties of the Zephyr bioharness device: a systematic review

Study

Criterion measure

Testing protocol

Agreement bias

95% LoA

Johnstone et al. (2012a). [24]

Polar-T31.

Treadmill – Walking and Running

−3.05

−32.20 – 32.20

Johnstone et al. (2012). [1]

Polar-T31

Walk-Jog-Run

0.02

−79.20 – 79.20

Kim et al. (2013). [25]

12-Lead ECG

Treadmill running

0.50

−15.30 – 16.30

Flanagan et al. (2014). [27]

5 Lead ECG

Cycle ergometer

3.00

−2.84 – 2.42

Dolezal et al. (2014). [28]

12 Lead ECG

Treadmill Walk

0.04

−0.05 – 0.12

Search (Crawl)

−0.01

− 0.12 – 0.10

Stairs

−0.13

−0.21 – − 0.04

Fast walk

0.03

−0.09 – 0.14

Smith et al. (2014). [29]

3 Lead ECG

Treadmill Walk

− 0.40

−0.70 – − 0.10

Search (Crawl)

−1.70

−3.1 – −0.40

Stairs

0.40

0.04–0.70

Gatti et al. (2014). [26]

5 Lead EKG

Sitting

−0.78

−5.10 – 3.60

Thoracic rotation.

−0.77

−5.40 – 3.90

Arm Lifting.

0.22

−7.70 – 8.10

Batting.

−2.51

−15.70 – 10.70

Weight moving.

−4.81

−22.50 – 12.90

Walking.

−1.68

−15.60 – 12.20

Rawstorn et al. (2015). [22]

12 Lead ECG

Treadmill Running

−0.30

−21.87 – 9.26

Treadmill, Cycle Ergometer and Activities of daily living; sweep and vacuum

1.10

−13.39 – 23.79

  1. ZB zephyr bioharness, 95% LoA limits of agreement