Table 5 Volume and thickness for harvested and unharvested tendons - two examiners taking two acquisitions on the same day

From: Reliability and methodology of quantitative assessment of harvested and unharvested patellar tendons of ACL injured athletes using ultrasound tissue characterization

Inter-rater reliability for acquisition and analysis - 2 examiners/ 2 acquisitions/ same day/ 2 analysis per tendon
6 Participants
Harvested (n = 11) and Unharvested (n = 10) patellar tendons
Areas of interest Volume (cm3) Thickness (cm)
Examiner1 (C.S.P.)
 Patellar Apex 1.1 ± 0.4 NA
 Proximal tendon 1.1 ± 0.4 NA
 Mid Tendon 1.0 ± 0.3 0.6 ± 0.3
 Distal Tendon 1.0 ± 0.4 NA
Examiner 2 (R.C.G.S.)
 Patellar Apex 0.9 ± 0.2 NA
 Proximal tendon 1.0 ± 0.2 NA
 Mid Tendon 1.0 ± 0.3 0.6 ± 0.2
 Distal Tendon 1.0 ± 0.3 NA
ICC (95% CI) 0.67 (0.52, −0.77) 0.97 (0.91, − 0.98)
SEM (ICC) % GrM 10.2 7.8
SEM (ICC) 0.1 0.05
  1. n = number of tendons assessed. Mean ± standard deviation
  2. ICC (95% CI) Intra-class coefficient of reliability (95% confidence interval), SEM % GrM Standard error of measurement as percentage of the grand mean, SEM Standard error of measurement